›› 2009, Vol. 8 ›› Issue (6): 305-308.

• 论著 • 上一篇    下一篇

血液透析患者中心静脉留置导管失功原因与处理

石宏斌 徐 航 梁冰茕 魏 乐   

  1. 广西中医学院附属瑞康医院肾内科
  • 收稿日期:2008-11-11 修回日期:1900-01-01 出版日期:2009-06-12 发布日期:2009-06-12
  • 通讯作者: 石宏斌

Causes and treatment of central venous catheter dysfunction in hemodialysis patients

SHI Hong-bin, XU Hang, LIANG Bin-qiong, WEI Le   

  1. Nephrology Department, The Ruikang Affiliated Hospital of Guangxi Traditional Chinese Medicine University, Nanning 530011, China
  • Received:2008-11-11 Revised:1900-01-01 Online:2009-06-12 Published:2009-06-12

摘要:

【摘要】目的 提高对中心静脉留置导管失功的认识和诊治水平,改善血液透析患者预后。方法 分析156例中心静脉置管的维持性血液透析患者。采用Seldinger技术置管,其中36例为长期双腔导管,120例为临时双腔导管。结果 导管失功有两大类:①导管通路功能障碍22例,原因以导管血栓形成为主(占59%),导管纤维外鞘形成是长期双腔导管功能障碍的另一个重要因素,在导管使用过程发生;于置管后即刻发生者多由导管打折、贴壁等技术因素引起;②导管相关性感染27例:糖尿病肾病(DN)患者感染发生率显著高于非DN患者;股静脉置管的感染发生率显著高于颈内静脉置管;临时导管感染发生率显著高于长期导管,且留置越久,临时导管感染发生率越高。结论 导管内血栓形成是两类导管通路障碍的主要原因,长期双腔导管发生障碍的重要原因还有纤维外鞘;临时导管留置时间延长及会增加感染的机会,影响感染的因素还有糖尿病和置管部位;需长期留置导管时,应用带涤纶套的导管可明显减少感染。

关键词: 血液透析, 中心静脉导管, 导管功能障碍, 导管相关性感染

Abstract:

【Abstract】 Objective To improve the recognition, treatment and prognosis of central venous catheter (CVC) dysfunction in hemodialysis patients. Methods We analyzed 156 maintenance hemodialysis patients with CVC, of which 36 used long-term indwelling cuffed catheter, and 120 used temporary CVC. Results Access dysfunction and catheter related infection were the two major causes leading to CVC dysfunction. Access dysfunction was found in 22 cases, of which 59% were thrombosis formation in CVC. Fibrinous sheath was another important cause of access dysfunction usually found on long-term indwelling cuffed catheter. Catheter related infection was frequently found in central vein catheterization through femoral vein and in patients complicated with diabetic nephropathy. The infection rate was higher in patients using temporary CVC than in those using long-term indwelling CVC. In temporary central vein catheterization, the longer the indwelling duration, the more the infection rate. Conclusion Thrombosis formation in catheter was the most common cause leading to CVC dysfunction. Fibrinous sheath was frequently found on long-term indwelling cuffed catheter. Catheter related infection is related to the patients with diabetic nephropathy, the catheterization site, the duration using temporary catheter. Catheter with a cuff was suitable for long-term indwelling CVC with less infection.

Key words: Central venous catheterization, Catheter dysfunction, Catheter related infection

中图分类号: