国产BLT-D800血滤机与进口Braun Dialog+血滤机的临床应用对比

李  琪   李晓颖   张庆华   张中琴   

  1. 550007 贵阳,1北京积水潭医院贵州医院血液净化中心
  • 收稿日期:2023-11-06 修回日期:2023-12-29 出版日期:2024-03-12 发布日期:2024-03-12
  • 通讯作者: 李晓颖 E-mail:896813035@qq.com

Comparison of the application of domestic BLT-D800 and imported Braun Dialog+ hemofiltration machines

LI Qi, LI Xiao-ying, ZHANG Qing-hua, ZHANG Zhong-qin   

  1. Blood Purification Center, Guizhou Hospital of Beijing Jishuitan Hospital, Guiyang, 550007, China
  • Received:2023-11-06 Revised:2023-12-29 Online:2024-03-12 Published:2024-03-12
  • Contact: 550007 贵阳,1北京积水潭医院贵州医院血液净化中心 E-mail:896813035@qq.com

摘要: 目的  研究Biolight D800(BLT-D800,广东宝莱特医用科技股份有限公司,中国)血滤机的性能。 方法 以Braun Dialog+(贝朗爱敦股份公司,德国)血滤机为对照,采用随机对照法研究国产BLT-D800血滤机性能。选择2022年8月北京积水潭医院贵州医院血液透析中心接受规律血液透析治疗的MHD患者120例,采用随机数字法分为2组(60例/组),每周3次血液透析(hemodialysis,HD),每月1次后稀释血液透析滤过(hemodiafiltration,HDF),对比HDF治疗时血液透析并发症、治疗效果等指标和设备配液情况。 结果  2组患者在HDF治疗过程中,低血压、高血压、肌肉痉挛、胸闷等不良反应发生率差异无统计学意义(χ2=0.178,0.934,0.167,0.164,P=0.677,0.358,0.666,0.873);BLT-D800组HDF治疗后尿素清除指数(Kt/V)、尿素下降率(URR)、血β2微球蛋白(β2 microglobulin,β2-MG)等指标的改善程度与Braun Dialog+组比较差异无统计学意义(t=-0.426,0.165,0.046,P=0.381,0.692,0.833);超滤误差对照组更高,2组间差异有统计学意义(t=-14.956,P<0.001)。 结论  国产BLT-D800血滤机性能稳定,能安全有效开展HDF治疗。

关键词: 血液透析滤过, 血滤机, 国产

Abstract: Objective To investigate the performance of domestic Biolight D800 hemofiltration machine (BLT-D800 Guangdong Biolight Meditech CO., China).  Method  This was a randomized controlled study, using the imported Braun Dialog+ hemofiltration machine (B. Braun Avitum AG, Germany) as the control. A total of 120 patients undergoing maintenance hemodialysis (MHD) at the Blood Purification Center, Guizhou Hospital of Beijing Jishuitan Hospital were recruited for this study. They were divided into 2 groups (n=60/group) by a random number table method. Patients in each group were treated with hemodialysis (HD) 3 times/week and post diluted hemodiafiltration (HDF) once/month. Dialysis complications, treatment effects, dialysate for the machine and other indicators were observed during HDF.  Result  The ratios of hypotension, hypertension, muscle cramp, and choking sensation in chest were similar between the two groups (χ2=0.178, 0.934, 0.167 and 0.164 respectively; P=0.677, 0.358, 0.666 and 0.873 respectively). After HDF treatment, Kt/V, urea reduction rate (URR) and serum β2 microglobulin (β2-MG) were also similar between the two groups (t=     -0.426, 0.165 and 0.046 respectively; P=0.381, 0.692 and 0.833 respectively). Ultrafiltration error was statistically higher in the control group (t=-14.956, P<0.001).  Conclusion  The domestic BLT-D800 hemofiltration machine is stable for blood purification performance, and is capable to carry out HDF safely and effectively.

Key words: Hemodiafiltration, Hemofiltration machine, Domestic